Development of the technology for ‘Diprifusor’ TCI systems

AnaesthesiaPublished 1 April 1998
John Gray, G. N. C. Kenny
Citations89
SJR quartileQ1
SJR score2.75
SNIP3.14

TL;DR

The ‘Diprifusor’ electronics module is a dual microprocessor component that can be incorporated into an infusion pump to provide this mode of administration and the safety aspects addressed by the design are described.

Abstract

The availability of fast-acting intravenous drugs for anaesthesia has encouraged anaesthetists to gain a practical understanding of the pharmacodynamics and pharmacokinetics of these drugs to optimise their administration. 'Diprifusor' technology uses pharmacokinetic modelling to control the infusion rate of a pump, providing the anaesthetist with direct control of the blood concentration rather than indirect control by the infusion rate. The 'Diprifusor' electronics module is a dual microprocessor component that can be incorporated into an infusion pump to provide this mode of administration. This paper describes the module's infusion control system, prefilled syringe recognition system and the safety aspects addressed by the design.

Keywords

MedicineNeuroscienceChemical Engineering