Determination of melatonin in blood plasma, using capillary gas chromatography and electron impact medium-resolution mass spectrometry

Journal of Mass SpectrometryPublished 1 March 1988
C. R. Lee, H. Esnaud
Citations16
SJR quartileQ3
SJR score0.43
SNIP0.66

TL;DR

It is shown that a general-purpose medium-resolution instrument gives adequate sensitivity in electron impact mode, when used with a capillary gas chromatographic column-switching device.

Abstract

Determination of the low concentrations (less than 10-100 pg ml-1) of melatonin in blood plasma by gas chromatography/mass spectrometry has previously required the use of negative ionization (electron capture) mass spectrometry, a technique particularly delicate to set up. We show that a general-purpose medium-resolution instrument gives adequate sensitivity in electron impact mode, when used with a capillary gas chromatographic column-switching device. A straightforward extraction procedure gives good specificity, and the limit of useful measurement is better than 10 pg per 1 ml sample.

Keywords

ChemistryBiochemistry, Genetics and Molecular Biology