Microplate coagulation assays.
Generate an AI Snapshot to get a quick, structured summary of this paper.
A concise AI-generated summary of the paper will appear here once you click Generate AI Snapshot.
TL;DR
These microplate-based blood coagulation assays take advantage of the microplate format and can be customized for specific purposes, such as quantifying purified factor IX or assessing protein C activity in plasma.
Abstract
This report describes the development of microplate-based blood coagulation assays. The assays require a kinetic microplate reader to follow changes in absorbance at 405 nm caused by the coagulating plasma. Procedures for performing prothrombin time and activated partial thromboplastin time tests are described with intra- and inter-assay variability of a few percentage points. The prothrombin time of normal plasma was 64.5 +/- 3.6 s, and the activated partial thromboplastin time was 69.8 +/- 3.2 s. Clotting times were prolonged when normal plasma was mixed with plasmas deficient in particular coagulation factors, as expected. These assays take advantage of the microplate format (small sample size and multiple simultaneous assays) and can be customized for specific purposes, such as quantifying purified factor IX or assessing protein C activity in plasma.