On the Scalability and Requirements of Whole Protein Mass Spectrometry

Analytical ChemistryPublished 11 July 2011
Philip D. Compton, Leonid Zamdborg, Paul M. Thomas, Neil L. Kelleher
Citations242
SJR quartileQ1
SJR score1.53
SNIP1.26

TL;DR

The model quantifies the challenge that top-down proteomics faces with respect to current MS instrumentation and projects that depletion of (13)C and (15)N isotopes can improve detection at high mass by only <2-fold at 100 kDa whereas the effect is up to 5-foldat 10 kDa.

Abstract

Top-down proteomics has improved over the past decade despite the significant challenges presented by the analysis of large protein ions. Here, the detection of these high mass species by electrospray-based mass spectrometry (MS) is examined from a theoretical perspective to understand the mass-dependent increases in the number of charge states, isotopic peaks, and interfering species present in typical protein mass spectra. Integrating these effects into a quantitative model captures the reduced ability to detect species over 25 kDa with the speed and sensitivity characteristic of proteomics based on <3 kDa peptide ions. The model quantifies the challenge that top-down proteomics faces with respect to current MS instrumentation and projects that depletion of (13)C and (15)N isotopes can improve detection at high mass by only <2-fold at 100 kDa whereas the effect is up to 5-fold at 10 kDa. Further, we find that supercharging electrosprayed proteins to the point of producing <5 charge states at high mass would improve detection by more than 20-fold.

Keywords

ChemistryEngineering