The Nellcor N‐101 pulse oximeter

AnaesthesiaPublished 1 March 1986Open access
J. A. TYTLER, H. F. SEELEY
Citations32
SJR quartileQ1
SJR score2.75
SNIP3.14
View PDF

TL;DR

The accuracy of the Nellcor N‐101 pulse oximeter has been evaluated in adult patients receiving general anaesthesia or intensive care and within the range tested (70–100 percent saturation of haemoglobin with oxygen) the readings were within I digit of the values obtained by in vitro measurement.

Abstract

The accuracy of the Nellcor N-101 pulse oximeter has been evaluated in adult patients receiving general anaesthesia or intensive care. Readings obtained noninvasively with this instrument were compared with measurements made on arterial blood using a Radiometer OSM2 oximeter. The pulse oximeter was easy to use and within the range tested (70-100 percent saturation of haemoglobin with oxygen) the readings were within I digit of the values obtained by in vitro measurement.