Pulse oximeters’ reliability in detecting hypoxemia and bradycardia: Comparison between a conventional and two new generation oximeters

Critical Care MedicinePublished 1 May 2000
Bettina Bohnhorst, Corinna Peter, Christian F. Poets
Citations115
SJR quartileQ1
SJR score2.12
SNIP2.15

TL;DR

Pulse oximeters are increasingly used for patient monitoring; however, they are traditionally very prone to motion artifact and one of the two new generation instruments investigated in this study missed 5.4% of hypoxemic episodes and 69% of bradycardias.

Abstract

OBJECTIVE: Pulse oximeters are increasingly used for patient monitoring; however, they are traditionally very prone to motion artifact. Newly developed instruments have lower false alarm rates. We wanted to know whether this is achieved at the expense of an increased proportion of false negative alarms such as missed or delayed identification of hypoxemia and/or bradycardia. DESIGN: Observational study. SETTING: Neonatal intensive care unit. PATIENTS: A total of 17 unsedated preterm infants (median gestational age at birth, 25 wks; range, 24-30 wks). INTERVENTION: Long-term recordings of transcutaneous partial pressure of oxygen (P(Tc)O2), heart rate, pulse oximeter saturation (SpO2), and pulse rate from a conventional oximeter and two new generation oximeters. MEASUREMENTS: Recordings were analyzed for episodes with P(Tc)O2 5 secs. Hypoxemia was considered identified if SpO2 had fallen to or =10 secs in 11 episodes; therefore, these episodes should have been identified by all three oximeters. None of these had been missed by the conventional oximeter, but 10 (5.4% of the total) were missed by one of the new generation instruments (Nellcor), and one (0.5%) was missed by the other (Masimo). Of 54 bradycardias, only 14 were identified by all three oximeters; 17 (32%) were missed by the conventional, 37 (69%) by the Nellcor, and 4 (7%) by the Masimo instrument. CONCLUSION: One of the two new generation instruments investigated in this study missed 5.4% of hypoxemic episodes and 69% of bradycardias. It thus appears that this instrument's reduced false alarm rate is achieved at the expense of an unreliable and/or delayed identification of hypoxemia and bradycardia. The other instrument identified both conditions equally as or more reliably than a conventional pulse oximeter.

Keywords

MedicineEngineering