Performance of the OraQuick® and Hema-Strip® rapid HIV antibody detection assays by non-laboratorians

Journal of Clinical VirologyPublished 5 February 2004
Timothy C. Granade, Bharat Parekh, Susan Phillips, J S McDougal
Citations44
SJR quartileQ1
SJR score1.19
SNIP0.90

TL;DR

Assessment of the ability of 99 individuals with no laboratory experience to conduct two RT, OraQuick and Hema-Strip, found the number of invalid tests was greatly reduced when participants received a demonstration.

Abstract

Rapid HIV antibody tests (RT) now permit HIV screening in settings where laboratory personnel may not be available. This study assessed the ability of 99 individuals with no laboratory experience to conduct two RT, OraQuick and Hema-Strip; these results were compared with those generated by laboratory professionals. All participants received written instructions and one-half also received a short demonstration. Error rates ranged from 2.1% to 4.6% with or without a demonstration. However, the number of invalid tests was greatly reduced when participants received a demonstration. Appropriate RT training for non-laboratorians and continued monitoring of HIV RT performance in non-laboratory settings is recommended.

Keywords

Immunology and MicrobiologyMedicine