Calibrator Design for the<i>COBE</i>Far‐Infrared Absolute Spectrophotometer (FIRAS)

The Astrophysical JournalPublished 20 February 1999Open access
John C. Mather, D. J. Fixsen, R. A. Shafer, Carol L. Mosier, D. T. Wilkinson
Citations498
View PDF

Abstract

The photometric errors of the external calibrator for the FIRAS instrument on the COBE are smaller than the measurement errors on the cosmic microwave background (CMBR) spectrum (typically 0.02 MJy/sr, 1 sigma), and smaller than 0.01% of the peak brightness of the CMBR. The calibrator is a re-entrant cone, shaped like a trumpet mute, made of Eccosorb iron-loaded epoxy. It fills the entire beam of the instrument and is the source of its accuracy. Its known errors are caused by reflections, temperature gradients, and leakage through the material and around the edge. Estimates and limits are given for all known error sources. Improvements in understanding the temperature measurements of the calibrator allow an improved CMBR temperature determination of 2.725 +/- 0.002 K.

Keywords

EngineeringPhysics and Astronomy