login

Laboratory validation of a diffusive sampler for the determination of glutaraldehyde in air

Journal of Chromatography APublished 1 August 1995
Roger Lindahl, Jan‐Olof Levin
Citations15
SJR quartileQ1
SJR score0.73
SNIP1.02

Abstract

Abstract A diffusive sampling method has been validated for determination of glutaraldehyde in air. The sampler consists of a filter impregnated with 2,4-dinitrophenylhydrazine, mounted in a polypropylene housing. The uptake rate was determined to be 11.8 ml/min, with a relative standard deviation of 13%. The effect of glutaraldehyde concentration, sampling time and relative humidity on uptake rate was undetectable or small. The temperature effect was significant, about 1.5%/°C. The samplers are stable for at least two weeks at 22°C. The detection limit is about 0.03 mg/m 3 for a 15-min sample.

Keywords

Immunology and MicrobiologyChemical Engineering