Complications of Lumbar Spinal Fusion with Transpedicular Instrumentation
Generate an AI Snapshot to get a quick, structured summary of this paper.
A concise AI-generated summary of the paper will appear here once you click Generate AI Snapshot.
TL;DR
This 5-year series included 486 patients who underwent 533 variable screw placement precedures for diseal, degenerative, and spondylolytic problems, and device modifications have reduced the indicence of screw breakage and nut loosening.
Abstract
Transpedicular screw fixation systems are coming into wide use as an adjunct to lumbar spinal fusion procedures. This 5-year series included 486 patients who underwent 533 variable screw placement procedures for discal, degenerative, and spondylolytic problems. The wound infection rate was 2.6%: 0.6% deep, 0.9% graft, and 1.1% superficial. The neural injury rate was 1.1% overall: 0.6% related to posterior lumbar interbody fusion and 0.6% related to instrumentation. Technical problems per procedure included 8.1% difficult screw placement, 5.6% nut loosening, and 4.3% screw breakage (1.1% per screw placed). Device modifications have reduced the incidence of screw breakage and nut loosening. No device-related neural injuries occurred in the last 333 procedures. With experience, the device can be applied safely without significantly increasing the risk of neural injury or morbidity.