Retrospective estimation and correction of physiological artifacts in fMRI by direct extraction of physiological activity from MR data

Magnetic Resonance in MedicinePublished 1 March 1996
Tuong Huu Le, Xiaoping Hu
Citations150
SJR quartileQ1
SJR score1.16
SNIP1.24

TL;DR

Experimental results illustrate that the physiological artifact reduction method provides effective compensation for physiology‐related signal fluctuations in functional MRI and performs as well as the retrospective technique using external physiological monitoring.

Abstract

A physiological artifact reduction method based on extracting respiratory motion and cardiac pulsation directly from functional MR data is described. In fast low angle shot (FLASH), respiratory cycles are derived utilizing the phase of the center of a navigator echo, in echo-planar imaging (EPI) from the phase of the center k-space point. Cardiac cycles are determined from projections obtained from the navigator echo (FLASH) and the center k-space line (EPI). Because direct extraction of physiological parameters eliminates the need for external monitoring, the method can be more readily implemented. Experimental results illustrate that the technique provides effective compensation for physiology-related signal fluctuations in functional MRI and performs as well as the retrospective technique using external physiological monitoring.