Intracranial Pressure Telemetry System Using Semicustom Integrated Circuits
Generate an AI Snapshot to get a quick, structured summary of this paper.
A concise AI-generated summary of the paper will appear here once you click Generate AI Snapshot.
TL;DR
A new generation of implantable, telemetric transmitters for intracranial pressure (ICP) measurements have been developed and long-term in vitro and in vivo pressure baseline stabilities of better than 1 and 2 mmHg per month, respectively, have been achieved.
Abstract
A new generation of implantable, telemetric transmitters for intracranial pressure (ICP) measurements have been developed. A unique technique used in packaging the silicon piezoresistive pt essure transducer provides excellent long-term stability. Pulse code modulation is used for data transmission over a radio frequency (RF) link. To minimize the component count, two semicustom, bipolar integrated circuits are used. The transmitter electronics are housed inside a 29 ×20 ×7 mm titanium package along with the pressure transducer and two lithium batteries. Even though the transmitter consumes less than 0.4 mW of power, it is turned on remotely via RF signal transduction only on demand in order to extend the lifetime of the batteries to years. The pressure input of the transmitter has a dynamic range of -100- +200 mmHg with a 0.3 mmHg resolution and a 1 mmHg accuracy. Long-term in vitro and in vivo pressure baseline stabilities of better than 1 and 2 mmHg per month, respectively, have been achieved.